Disclaimer: The information available on the Sarvodaya Healthcare website, including but not limited to Centres of Excellence, Departments, Treatments, Health Information, Blogs, FAQs, Videos, Graphics, Illustrations, Rare Cases, and other related content, is provided solely for general informational and educational purposes.
The information published on this website is not intended to be a substitute for professional medical advice, consultation, diagnosis, treatment, or prescription. It should not be relied upon for diagnosing or treating any medical condition or for making decisions regarding your health or medical care.
Every individual’s medical condition is unique. We strongly recommend consulting a qualified healthcare professional for appropriate medical assessment, diagnosis, treatment, and advice based on your specific health needs and circumstances.
Abstract
Context: Late-night salivary cortisol (LNSC) is a simple and reliable screening test for Cushing syndrome (CS). With improved analytical performance of the current second- generation electrochemiluminescence immunoassay (ECLIA; Elecsys Cortisol-II; Roche Diagnostics), there is a need to revisit the LNSC cut-offs, especially in a South-Asian population.
Objective: To derive LNSC cut-offs for diagnosis of CS using second-generation ECLIA kits.
Design: Diagnostic accuracy study.
Methods: We prospectively recruited 155 controls aged 18–60 years, including, normal-weight (body mass index [BMI] < 25 kg/m2 and no hypertension or diabetes [n = 53]) and overweight/obese (BMI 25–30 kg/m2 and hypertension and/or
diabetes [n = 52] or BMI = 30 kg/m2 with/without comorbidities [n = 50]) participants. All participants submitted LNSC samples collected at home; overweight/ obese controls additionally underwent dexamethasone suppression test to exclude CS. We also reviewed records of adults with endogenous CS (cases, n = 92) and a valid LNSC result using the same method.
Results: The 95th percentile for LNSC in controls was 6.76 nmol/L. The mean ± SD LNSC levels were 40.47 ± 49.63 nmol/L in cases and 3.37 ± 1.18 nmol/L in controls (p < 0.001). Receiver operating characteristic (ROC) analysis showed excellent diagnostic performance of LNSC for CS, with area under curves (AUCs) of 0.994 (cases vs. all controls) and 0.993 (cases vs. overweight/obese controls), respectively. The best diagnostic performance was achieved at cut-offs =6.73 nmol/L (sensitivity: 97.8%, specificity: 94.8%) and =7.26 nmol/L (sensitivity: 97.8%, specificity: 95.1%), respectively.
Conclusions: LNSC measured using second-generation ECLIA demonstrated high diagnostic accuracy for CS. Based on this study, we propose a LNSC cutoff =6.73 nmol/L to diagnose CS.